16 GOSSEN Foto- und Lichtmeßtechnik GmbH
3.4.2 Measuring correction values
Correction values can also be directly measured.
However, measurement does require constant li-
ght.
Depression of the measuring button initiates a
reference measurement that is identified on the
digital display by rF --.
Insert the light-reducing filter (e.g. neutral density
filter) in the optical path, and press the measu-
ring button.
The light reducing factor in stops, and the exten-
sion factor, automatically appear on the display.
Pressing the function buttons transfers the
correction values to the memory of the
SIXTOMAT flash.
The box around still appears when a corre-
tion value was entered.
The COR value is now automatically considered
in all subsequent measuring functions.
COR